Particulars
The Specifications of HX-801B
Model: |
HX-801B
|
Setting Model |
Flow Rate
|
Pumping type: |
Peristaltic finger pump
|
IV set requirement: |
Standard IV set (10,15,20drops/ml),or micro drip IV set (60drops/ml) with 3.4-4.5mm tubing
|
Flow rate setting range: |
1.0ml/h - 999.9ml/h
|
Drip rate setting range: |
IV set of 10 drops/ml:1.0-166.6drops/min
IV set of 15 drops/ml:1.0-249.9drops/min
IV set of 20 drops/ml:1.0-333.3drops/min
IV set of 60drops/ml:1.0-150.0drops/min
|
Flow rate increment |
0.1ml/h
|
Flow rate accuracy: |
+/- 5% (standard PVC set must first be adjusted)
|
Drip rate accuracy: |
+/- 2%
|
Volume limit setting range |
1-9999 ml
|
Operational condition: |
Temperature:5- 40; Relative humidity: ?90%
|
Volume infused display : |
0-9999 ml
|
Finger Cassette |
Detachable
|
Purge rate: |
600 ml/h
|
KVO(keep vein open): |
1ml/ h
|
Air detection method : |
Ultrasonic, Check range 50~1000ul ,8 levels can be adjusted
|
Drip detection method: |
Infrared photo-electric beam
|
Display: |
Color LCD with high backlight
|
Occlusion pressure : |
Special prssure sensor,40kPa-140kPa ,8 levels occlusion sensitivity
|
Alarms: |
Infusion Monitoring:
Air in tube,door open,occlusion, pause,infusion complete,empty,
abnormal drips,leakage,tube slope etc.
Diagnosis malfunctions:
Low battery,damage battery,abnormal drip detector,abnormal peristaltic finger pump,reversal of motor,abonormity of ultrasonic sensor,abnormal of motor,abnormal of alarm sound,pause of the power supply
|
Keyboard |
PET Film keyboard
|
Door Buckle |
Double-buckle
|
Adjustment |
Automatical, manual( Optional)
|
Power source: |
100V-240V , 50Hz/60Hz
|
Inner Battery: |
Rechargeable Ni- hydrogen battery, DC12V , 2000mAh
|
Discharge Time: |
> 4 hours
|
Power consumption: |
<25VA with power supply; <5W with battery supply
|
Weight: |
2.1kg
|
Classification: |
Class II; Internal power supply; type CF; Portable ; IPX4
|
Dimensions: |
151mm*150mm*235mm
|
User range: |
Venous infusion,Vascular expansion solution,Anticancer drugs,Accelerative procreate drugs,Nutrient liquid and all of liquid medicine, mini infusions and rapid infusions.
|
|
|